GD3128 Ikizamini cyo Kurwanya Kurwanya
Ikizamini cya GD3128 Ikizamini cya Insulation Resistance Tester gifite ibikorwa byiza byo gupima ibipimo bitandukanye byo kurwanya insulasiyo hamwe nubushobozi buhebuje bwo kurwanya-interineti, bushobora gukoreshwa mugupima imbaraga zo gukumira insuline nini y’ibikoresho by’amashanyarazi nini cyane hamwe n’umurongo wohereza mu bidukikije bikabije nka insimburangingo.Igikoresho gikoresha ibara ryukuri ryo gukoraho, amakuru yikizamini hamwe nimbaraga za batiri zerekanwa kuri ecran imwe iyo urebye, kandi guhuza knob switch nurufunguzo bituma imikorere yoroshye cyane.Umukoresha ntabwo akeneye gufata mu mutwe uburyo bwo gukora.Kanda "GUFASHA" kuri ecran kugirango ubone uburyo bwo gukora, bushobora kuyobora neza umukoresha gukoresha igikoresho.Igikoresho gifite imikorere yo gusohora byihuse, gishobora gusohora voltage yikintu cyageragejwe ako kanya nta yandi mashanyarazi asohoka.Umuyoboro wa voltage wigikoresho urashobora guhita ukurikirana voltage nzima yikintu cyageragejwe.Niba voltage irenze 36V, ikizamini kizahita kibuzwa, gishobora kurinda neza igikoresho nabakora.
Igikoresho gikoresha ubushobozi bunini bwa lithium yamashanyarazi.Igihe cyikizamini cyigikoresho gihita cyandika igihe cyikizamini kandi kibika ibisubizo byikizamini nitariki nigihe.Mugukoraho ecran irashobora kubona byoroshye amakuru yanditse mumateka.Imigaragarire ya USB yitaruye rwose irashobora kohereza amakuru yikizamini kuri PC.Igikoresho gifite umurongo wikizamini hamwe nubushakashatsi bubiri hamwe ningingo ikingira, kandi umuyaga mwinshi wa voltage ufite ibikoresho bisimburwa na alligator clip hamwe na hook, bishobora guhuza ahantu hatandukanye.Igikoresho gifata ibyuma bibiri byubatswe, kandi urwego rwo kurinda agasanduku ko hanze ni IP65, rushobora gukumira igitero cy’amazi n’umukungugu mugihe cyo gutwara no kubika, kandi bikarinda neza igikoresho kugongana.
Urukurikirane rwa GD3128 rufite intera nini, ikemurwa cyane, imikorere yoroshye, kwinangira no kuramba, ubunyangamugayo no kwizerwa, imikorere ihamye, hamwe nubushobozi bukomeye bwo kurwanya kwivanga.Igikoresho gifite ikizamini cyo kurwanya insulasiyo (IR);Ikizamini cya polarisiyonike (PI);ikigereranyo cyo kwinjiza (DAR);ikizamini cya ramp (RAMP);akayunguruzo (FR, 10S, 20S, 30S, 40S amahitamo ane);ikizamini cya voltage (V);ikizamini cya capacitance (uF);ikizamini kiriho (nA) nibindi bikorwa.Igikoresho kirakwiriye kwipimisha irwanya ibikoresho byamashanyarazi, imirongo yohereza hamwe nibindi bikoresho binini munsi yubushobozi bunini, voltage nini na induction ikomeye.Nibikoresho bisanzwe bikoreshwa kandi byingirakamaro mumashanyarazi, itumanaho, meteorologiya, icyumba cya mudasobwa, umurima wa peteroli, gushyiramo amashanyarazi no kuyitaho, hamwe nishami rishinzwe gutanga amashanyarazi yinganda n’amabuye y'agaciro.
●Umubare ntarengwa wumuzunguruko ni 7mA.
●Ikizamini cyo kurwanya insulasiyo (IR), ikizamini cyerekana polarisiyasi (PI), ikigereranyo cyo kwinjiza (DAR).
●Uburyo bwikizamini cya Ramp (RAMP), uburyo bwo kugerageza (FR), ikizamini cya capacitance (uF)
●Imikorere ya voltage imikorere, ihita ikurikirana voltage yishyuwe yikintu cyageragejwe.
●Imikorere yigihe cyigihe, ihita yandika igihe cyibizamini.
●Igikorwa cyo gusohora mu buryo bwikora, mu buryo bwikora kandi bwihuse kurekura amafaranga yikintu cyikizamini nyuma yikizamini, nta muyoboro usohoka hanze.
●Imikorere yo guhagarika byikora.
●Ibara rinini ryo gukoraho.
●Igice cya kabiri cyibishishwa byubaka, birakomeye kandi biramba, urwego rwo kurinda IP65.
●Ubushobozi buke bwa lithium yamashanyarazi.
●Imikorere yo kubika, irashobora guhita ibika amatsinda 1000 yamakuru yigihe-cyibizamini hamwe nitariki yikizamini nigihe nigihe cyo kugerageza.
●Gukuramo imikorere, ifite ibikoresho bya USB, amakuru yamakuru arashobora koherezwa kuri mudasobwa kugirango isesengure ryibarurishamibare.
Icyitegererezo cyibicuruzwa | Umuvuduko ukabije | Urwego rwo Kurwanya Kurwanya | Inzira ngufi |
GD3128A | 250V,500V,1kV,2.5kV,5kV | 0.50MΩ ~ 10.0TΩ | ≥7mA |
GD3128B | 250V,500V,1kV,2.5kV,5kV, 10kV | 0.50MΩ ~ 20.0TΩ | ≥7mA |
GD3128C | 250V,500V,1kV,2.5kV,5kV, 10kV, 15kV | 0.50MΩ ~ 30.0TΩ | ≥7mA |
Urwego rwo Kurwanya Kurwanya no Kwirinda
Umuvuduko ukabije | Umujinya wo Kurwanya | Ukuri | Ongera wibuke |
250V | 0.50MΩ~50.0GΩ | ± 5% rdg ± 3dgt |
|
50.0GΩ~500GΩ | ± 20% rdg ± 3dgt |
| |
500V | 1.00MΩ~100GΩ | ± 5% rdg ± 3dgt |
|
100GΩ~1.00TΩ | ± 20% rdg ± 3dgt |
| |
1000V | 1.00MΩ~200GΩ | ± 5% rdg ± 3dgt |
|
200GΩ~2.00TΩ | ± 20% rdg ± 3dgt |
| |
2500V | 1.00MΩ~500GΩ | ± 5% rdg ± 3dgt |
|
500GΩ~3.00TΩ | ± 20% rdg ± 3dgt |
| |
5000V | 1.00MΩ~1.00TΩ | ± 5% rdg ± 3dgt |
|
1.00TΩ~10.0TΩ | ± 20% rdg ± 3dgt |
| |
10KV | 1.00MΩ~2.00TΩ | ± 5% rdg ± 3dgt |
|
2.00TΩ~20.0TΩ | ± 20% rdg ± 3dgt |
| |
15KV | 1.00MΩ~3.00TΩ | ± 5% rdg ± 3dgt |
|
3.00TΩ~30.0TΩ | ± 20% rdg ± 3dgt |
1 TΩ (Tera ohm) = 1000GΩ = 1012Ω
1 GΩ (Giga ohm) = 1000MΩ = 109Ω
1 MΩ (Mega ohm) = 1000KΩ = 106Ω
Imikorere | Ibipimo byo kurwanya insulasiyo (IR);ibipimo byerekana ibipimo (PI);igipimo cyo kwinjiza (DAR);uburyo bwo kugerageza ramp (RAMP);muyunguruziimikorere(10S, 20S, 30S, 40S);vgupima oltage (V);capacitancegupima(uF);ibipimo bigezweho(nA) |
Amashanyarazi | Bateri ya lithium ishobora kwishyurwa 6.2Ah |
Umuvuduko ukabije | GD3128A: 250V, 500V, 1kV, 2.5kV, 5kV GD3128B: 250V, 500V, 1kV, 2.5kV, 5kV, 10kV GD3128C: 250V, 500V, 1kV, 2.5kV, 5kV, 10kV, 15kV |
Ibisohoka Umuvuduko Ukwiye | (5%~10%) ± 10V |
Ikizamini cyo Kurwanya Kurwanya | GD3128: 0.50MΩ~10.0TΩ GD3128B: 0.50MΩ~20.0TΩ GD3128C: 0.50MΩ~30.0TΩ |
Ibisohoka Bigufi Inzira Zigezweho | ≥7mA |
Ikizamini cya Polarisiyasi | Yego |
Ikigereranyo cya Absorption Ikigereranyo | Yego |
Uburyo bwo Kugerageza | Gerageza mu ntambwe ya 10% ya voltage yateganijwe kugeza voltage yagenwe. |
Akayunguruzo Ikizamini | 10S, 20S, 30S, 40S byatoranijwe. |
Ikizamini cya voltage | Urwego: AC / DC: 0V~1000V; Ukuri: ± 5% rdg ± 3V |
Ikizamini cyubushobozi | Urwego: 10nF~25uF; Ukuri: ± 10% rdg ± 10nF |
Ikizamini Cyubu | Urwego: 0.01nA~7mA; Ukuri: ± 5% rdg ± 0.5nA |
Umugenzuzi w'amashanyarazi | Kurikirana voltage yatested ikintu, kandi ukurikirane uko ibintu byasohotse nyuma yikizamini, bibujijwe kugerageza mugihe voltage irenze 36V, kurinda ibikoresho nuwabikora. |
Ikizamini | Ikizamini cyikora cyikora igihe, igiheingintera: 0s ~ 9999s |
Imikorere yo kubika | Mu buryo bwikora ubike amakuru yikizamini hamwe nitariki yikizamini nigihe, byose hamwe mumatsinda 1000 |
Kuramo Imikorere | Kuramo amakuru yabitswe kuri mudasobwa ukoresheje USB itumanaho. |
Amashanyarazi ya Batiri | Hamwe na bateri yerekana ingufu, mugihe bateri ya voltage nkeya izibutsakwishyuzabateri. |
Guhagarika byikora | Nyuma yiminota 15 gutangira bizahita bifunga nta gikorwa. |
Igipimo | 280mm × 260mm × 160mm |
Ibiro | 4900g (shyiramo bateri) |
Umugozi wikizamini | Ibara ritukura hejuru ya voltage igerageza wire 1pc (hamwe naa clip ya alligator na hook), ibara ryicyatsi kibisi 1pc, ibara ryirabura ryirabura wire 1pc |
Urwego rwo Kurinda | Funga urubanza IP65, fungura urubanza IP40 |
Ibidukikije bikora | -20 ℃~50 ℃; 80% rh |
Kubika Ibidukikije | -25 ℃~65 ℃; 80% rh |
Kurwanya Kurwanya | 50MΩ(1000V) (hagati yikizamini cyizunguruka nigikonoshwa) |
Ihangane na voltage | AC 3kV50Hz 1min(hagati yikizamini cyizunguruka nigikonoshwa) |
Umutekano | IEC61010-1, IEC61326-1 |