GDYB-S20 Sisitemu Yicyiciro Cyicyiciro cya gatatu
Urutonde rwa GDYB Ibice bitatu bingana Byinshi Bikora Imikorere Yumuriro Wamashanyarazi Ikigereranyo cyateguwe neza mugisekuru cya gatatu cya elegitoroniki ishobora gutegurwa ibyiciro bitatu byamashanyarazi ya metero nyinshi sisitemu yo kwipimisha ukurikije ibipimo ngenderwaho bigezweho byigihugu.Yemeza 0,01% murwego rwo hejuru 1: 1 ikoranabuhanga rihindura, kugirango igere kuri voltage yimbere ihuza shunt metero eshatu.Irakwiriye kwipimisha metero eshatu zingufu zamashanyarazi zingana na 0.2% no munsi (zirimo metero yumuriro wa sinusoidal).Ifite isura nziza, imikorere yuzuye nibikorwa byiza.
GB / T 11150-2001 Ibikoresho byo gupima metero z'amashanyarazi (neq IEC 60736)
JJG 307-2006 Ibipimo by'amashanyarazi yo gupima Ingufu-z'amashanyarazi
JJG 596-1999 Ibipimo by'ingufu z'amashanyarazi hamwe na Electronics
JJG 597-2005 Ibikoresho byo kugenzura kubipimo by'amashanyarazi ya AC
DL / T 614-1997 Imikorere ya metero nyinshi
DL / T 614-2007 Ibikoresho byinshi bipima amashanyarazi (AC)
Ibidukikije
Gukoresha ubushyuhe bwibidukikije: 23 ℃ ± 2 ℃.
Gukoresha ubuhehere bwibidukikije: 45 % ~ 75 %.
Amashanyarazi
Icyiciro kimwe: AC220V ± 33V.Inshuro: 50 ± 2.5Hz
Ubwoko Oya.: GDYB-S20 Sisitemu Yicyiciro Cyicyiciro Cyamashanyarazi.
Sisitemu yukuri: 0.05%.0.1%.
Uburebure bwa metero isanzwe: 0,05% Ibyiciro bitatu byimikorere myinshi ya metero isanzwe.
Umuvuduko & Ibipimo bigezweho:
Intambwe ya voltage isohoka: 57.7V / 100V / 220V / 380V.
Ibisohoka bigezweho: 0.005A, 0.025A, 0.1A, 0.25A, 0.5A, 1A, 2.5A, 5A, 10A, 25A, 50A, 100A.
Ibisohoka byibuze bigezweho: 1mA.Ukuri: ≤5%.
Ubushobozi bwo gutwara ibintu (metero 20)
Umuvuduko w'amashanyarazi: 800VA.
Ibisohoka ubu: 2500VA.
Umuvuduko & Guhindura
Urutonde rushobora guhinduka: (0% ~ 120%) x igipimo cyo gupima.
Guhindura neza: 10%, 1%, 0.1%, 0.01%.
Guhindura icyiciro:
Urwego rushobora guhinduka: 0 ° ~ 360 °.
Ubwiza bushobora guhinduka: 10 °, 1 °, 0.1 °, 0.01 °.
Guhindura inshuro:
Urutonde rushobora guhinduka: 45Hz ~ 65Hz.
Guhindura neza: 1Hz, 0.1Hz, 0.01Hz.
Imbaraga zisohoka zihamye: ≤0.05% / 120S.
Ibisohoka voltage & kugoreka ibintu:
Umuvuduko: ≤0.5%.
Ibiriho: ≤0.5%.
Igenamiterere rihuza:
Umubare uhuza: inshuro 2 ~ 21.
Ibirimo bihuye: inshuro 2 ~ 21 ≤40% (ugereranije numuhengeri wibanze).
voltage inshuro 2 ~ 21 ≤40% (ugereranije numuhengeri wibanze).
Icyiciro gihuza: 0 ° ~ 359.99 °.
Igikoresho cyo gukurikirana:
Imetero isanzwe yerekana neza: Umuvuduko: 0.01%.
Ibiriho: 0.01%.
Imbaraga: 0.01%.
Icyiciro: 0.01.
Inshuro: 0.01.
Ironderero 1: 1 ibipimo bya transformateur:
Ukuri: 0.01%.
Imbaraga zisohoka: 60VA.
Ikigereranyo cyo gupima ubu: 1mA ~ 120A (1: 1).
Icyerekezo nyamukuru
a) Ukuri: 0.01% (COS∮ = 0.5C… 1… 0.5L).
b) Ikigereranyo cyagenwe: 50Hz / 60Hz ± 5%.
c) Ibanze ryibanze: 3 × 0.001-120A.
d) Icyiciro cya kabiri: 3 × 0.001-120A.
e) Ibisohoka bisohoka (buri cyiciro):
0.5W (ibisohoka: 0.001-1A).
0.7V igabanijwe nagaciro keza (ibisohoka: 1-120A).
f) Ibisohoka bitwara voltage: Byinshi.0.5V (ibisohoka muri iki gihe Max. 120A).
Ibipimo byahinduwe muri iki gihe: 265 x 200 x 240 mm (L x W x H).
Uburemere: 12kg.
Igipimo:
Inkomoko yo kugenzura: 610 x 800 x 1920mm (L x W x H).
Ikigereranyo cya metero: 2350 x 750 x 1850mm (L x W x H).