GDYM-3F Igendanwa Multi-imikorere yingufu za metero Calibrator
●Mugihe cyo kubara ibipimo byose no kwerekana kuri interineti harimo umuraba wuzuye, umuraba wibanze hamwe nu murongo uhuza icyarimwe mugihe kimwe cyerekana U, I, P, Q, VA, impande zinguni zerekeranye numuhengeri wuzuye hamwe numuhengeri wibanze kimwe na synthesis ya I, U, P, Q. PF, imiyoboro ihuza PF.
●Uburyo bwo kugenzura burimo ubwoko 7 bwinzira zuzuye zuzuye, imbaraga zuzuye zuzuye, imbaraga zuzuye zigaragara, umuraba wibanze ukora, imbaraga zifatika zifatika, imbaraga zifatika zigaragara kandi zihuza, zishobora kugenzura imbaraga zifatika za metero yuzuye, ibyingenzi metero, metero yumurongo wa metero hamwe nimbaraga zifatika zumurongo wuzuye hamwe na metero yumurambararo, birashobora kandi kugenzura ukuri kugaragara kwingufu za mashanyarazi.
●Guhitamo uko bishakiye kimwe cyangwa byinshi bihuza kugirango bibare guhuza ibice U, I, P, Q, VA nijanisha ryayo kimwe nimbaraga zihuza imbaraga zingufu zamashanyarazi.
●Icyarimwe icyarimwe gusesengura ibice no gupima agaciro ka buri gihe U, I, P, Q, VA yibyiciro bitatu byuzuzanya kandi bikerekana inshuro 51 + PH / -PH nijanisha ryumuraba wibanze kugirango umenye ingaruka zibyiza nibibi kuri gupima.
●Imbaraga zuzuye zifata imbaraga ni algebraic igiteranyo cyingufu zifatika zumuraba wibanze hamwe numuhengeri wa buri guhuza.Kugirango wirinde icyiciro cya 90 cyahinduwe mugihe cyashize no mubyerekeranye na harmonic wave reactive power test test.Turashobora gupima ikirango hamwe na metero yingufu zamashanyarazi zishobora kumenyekana no guhitamo umubare wa buri muhengeri uhuza.
●Icyarimwe gerageza impiswi ebyiri zamenyekanye, kandi burigihe, uburyo bwakazi numubare wabonetse mugihe kimwe birashobora kuba bitandukanye rwose.
●Kurenza inshuro 2,4, ingingo zose zapimwe zirashobora kwemeza neza ko 0,05% ari ukuri kandi ikanemeza ibipimo ngenderwaho bya CT icyiciro cya kabiri cyagenwe mugihe ari 1A.
●Inshuro 500 urwego rugezweho rwemeza ibipimo byo gupima, inshuro 1000 zerekana ibyerekanwe, byibuze 1mA itangira ikigezweho.Mugihe kimwe cyo gutanga amashanyarazi menshi, hapimwe ibikoresho byinshi-voltage, kandi insinga irashobora kumenyekana nta mutwaro.
●Imikorere yo gukusanya ingufu z'amashanyarazi irashobora gupima metero y'amashanyarazi idafite umusaruro.
●Erekana igishushanyo cya vector, irashobora kwerekana insinga eshatu za 48 insinga zitari zo hamwe ninsinga enye za 96 insinga zitari zo kimwe na CT ya kabiri.
●Kwemeza 7 inch 16: 9 urwego rwinganda nubugari bwamabara ya ecran ya TFT, buri kintu cyerekana imikorere yibikoresho byerekanwe kandi bisobanutse kuri ecran icyarimwe nta ecran ya ecran, yemerera abakoresha gukoresha iyo urebye.
●Igikoresho kirashobora kwigana ibintu bitandukanye byerekana insinga hanyuma ukabona ibisubizo bijyanye na wiring na coeffisente yo gukosora mubiro utabonye voltage cyangwa ibimenyetso byubu.Iyi mikorere irashobora gukoreshwa nkigikoresho cyo kugenzura umurongo wo guhugura kugirango utezimbere umurongo wo kugenzura umurongo w'abakozi bo murwego.
●Igikoresho cyimbere gishobora gutanga ibimenyetso byukuri bya 0.05-5A hanyuma ukongerwaho muri 5A clamp kugirango ugenzure amakosa ya metero ya clamp.Gutekereza kuri 5A clamp ifite ikibazo cyo guhindura amakosa nyuma yo gukoreshwa igihe kirekire.Gushoboza kumenya ikibazo cyo gufunga metero ya clamp umwanya uwariwo wose, ibutsa abakozi bireba mugihe cyoza urwasaya.Iyo usimbuye clamp ya 5A, ikosa rirashobora gukosorwa kuri 0.2.
●Mu buryo bwikora winjire muri metero yingufu (imikorere idahwitse)
Shyigikira barcode scaneri hanyuma uhite winjiza kode ya metero yingufu.
●Koresha 485 cyangwa infrarafoto yumutwe kugirango usome metero nyinshi zingufu za metero, impinga, igorofa, ikibaya, igiteranyo gikora kandi gikora, soma buri gihe nigihe.Shyigikira amategeko yigihugu ya DL645, DL645-1997 na DL645-2007.(imikorere idahwitse)
●Yubatswe muri 2G disiki ya elegitoroniki.
●Irashobora kuba ifite imashini icapura ya santimetero 3, ikoreshwa na bateri, kandi igahindura mu buryo butaziguye ishusho yerekana ibara ryerekana ibara ry'umukara n'umweru iyo icapwa. (Bihitamo)
●Ukoresheje metero ya clamp (5A, 100A, 500A, 1000A) kugirango upime igipimo nikosa ryikigereranyo cya voltage ntoya.(100A, 500A, 1000A clamp irahitamo)
●Erekana imbaraga za voltage nubu ibyiciro bitatu A, B, C.
●Imbaraga zakazi zinjiza ni 55V-600V.Ubusanzwe ibyiciro bitatu bya voltage yinjiza inganda sock ni 600V.Inganda zihindura inganda zihuza umujyi amashanyarazi na voltage yicyiciro cya gatatu nkimbaraga zakazi.
●Ibipimo byose bigomba kwinjizwa kurubuga no gukuramo amakuru birashobora gusobanurwa kubuntu kubakoresha.Irashobora gukuramo amakuru yabakoresha 5000.
●Mugaragaza ubu irashobora kubikwa kuri U disiki.
●Hamwe nubushuhe bwo gupima ubushyuhe nubushuhe (ukeneye gushiraho ubushyuhe nubushuhe).
●Hamwe na USB interineti kandi ihuza USB flash ya disiki kugirango ikore umwanya munini wo gukoporora amakuru.
●Kora kuri ecran hanyuma ukande urufunguzo rukora, clavier ukoresheje mudasobwa ya mudasobwa ya gahunda.Mwandikisho ya PC ifite USB interineti nayo irahari.
●Igihe hamwe na PC imashini ukoresheje interineti ya USB.
●Mudasobwa ya tablet irashobora kugenzura neza igikoresho gifite interineti ikora nkigikoresho.
●Umuvuduko wo gupima amashanyarazi hamwe nu guhinduranya amashanyarazi mubyiciro byinganda birashobora gukoreshwa munsi ya 600V (Imbaraga zisanzwe zikoreshwa ni 250V, zikoreshwa murwego rwo hejuru kuri 380V, zishobora guhungabanya umutekano).
●Igishushanyo mbonera cyumutekano.Imiyoboro nyamukuru, voltage ipima sock, iyinjiza ryubu, clamp yinjiza irihariye, hamwe na voltage ihagaze ni 4kV.
●Kurinda ingufu z'amashanyarazi.Mugihe winjije voltage nini izahita itangira kurinda no kwerekana ubutumwa bwihuse kubikoresho, ikosa rimaze gukosorwa, kanda kuri interineti kugirango usubire kurinda.
●Clamp ya 5A ifite imikorere yo kwikosora: ibimenyetso bya 5A bigezweho byakozwe imbere imbere nigikoresho cyongewe kuri 5A clamp kugirango bapime ikosa rya metero ya clamp.
●Imbaraga 5 za ARM chip zikoreshwa, ubushobozi bwo kubara burenze 1GIPS, kandi ibipimo byose bya 51 bihuza hamwe numuhengeri wuzuye ubarwa icyarimwe.
●Ukoresheje sisitemu ya Windows CE hamwe na sisitemu ikuze kandi yukuri, imikorere yayo irahamye kandi yizewe.
Umuvuduko | Ingano yamakosa | 25-600V | |||
Gutangira gupima agaciro | 0.5V | ||||
Agaciro kagereranijwe | 440V | ||||
Amakosa | ± 0,05% | ||||
Ibiriho | Terminal | IkigereranyoIe | 5A | ||
Urwego | 0.025A-12A | ||||
Clamp | IkigereranyoIe | 5A, 100A, 500A,1000A(bidashoboka) | |||
Urwego | 0.5% Ie - 2.4Ie | ||||
Gutangira | 0.02% Ie | ||||
Iyinjiza inshuro | 45-65Hz | ||||
Ikosa ryo gupima inshuro | ± 0.01Hz | ||||
Icyiciro cyo gupima icyiciro | -180 ° - 180 ° | ||||
Icyiza.Ikosa ry'icyiciro | ± 0.1 ° | ||||
Ijanisha rya Harmonic, ibipimo byamashanyarazi bihuza, hamwe nigihe cyo gupima ingufu zamashanyarazi | 2-51ibihe | ||||
Impanuka ihoraho (Icyitonderwa:Ieni Ikigereranyo) | Ibisohoka bike(Ibikoresho 3) | 5A | 3600P, 36000P, 360000P / kWh (kvarh, kVAh) | ||
abandi | 3600P, 36000P, 360000P * (5 / Ie) P / kWt (kvarh, kVAh) | ||||
Imbere murwego rwo hejuru | 5A | 1.8x109P / kWt | |||
Ibindi bikoresho bigezweho | 1.8x109* (5 / Ie) P / kWt | ||||
Umubare ntarengwa winjiza inshuro | 50kHz | ||||
5Aterminal | Umuhengeri wuzuyekWh, P, U, I | ± 0,05% | |||
Umuhengeri wuzuye Kvarh, Q, VA, VAh | ± 0.2% | ||||
HarmonicumurabaU., I, P, Q, S | ± 0.2% H.(H = 1 + 0.01k K = 2-51) | ||||
5Aclamp | Umuhengeri wuzuyekWh, P, U, I | ± 0.2% | |||
Umuhengeri wuzuye Kvarh, Q, VA, VAh | ± 0.2% | ||||
HarmonicumurabaU., I, P, Q, S | ± 0.5% H.(H = 1 + 0.01k K = 2-51) | ||||
Kwisuzumisha ubungubu ibyasohotse bisohoka | 0.05A, 0.1A, 0.25A, 0.5A, 1A, 2.5A, 5A | ||||
Kwigengakugenzuraikosa (itandukaniro n'umurongo) | ± 0.1% | ||||
Amashanyarazi manini | Umuhengeri wuzuyekWh, P, U, I | ± 0.5% | |||
Umuhengeri wuzuye Kvarh, Q, VA, VAh | ± 0.5% | ||||
HarmonicumurabaU., I, P, Q, S | ± 1.0% H.(H = 1 + 0.01k K = 2-51) | ||||
Abandi | Ibipimo | ± 0.5% | |||
Umuvuduko w'amashanyarazi | <± 0.01% | ||||
InshuroIngaruka | <± 0.01% | ||||
UbushyuheIngaruka | <± 0.005% / ℃ | ||||
Amasaha 24 ahindura ikosa | <± 0.01% | ||||
HarmonicIngaruka | <± 0.01% | ||||
Bateri yumuriro | 11.1V 3000mAh | ||||
Ihangane na voltage (Ibipimo bya voltage bigezweho hamwe n'amashanyarazi hamwe na terefone ntoya) | 4.0 kV | ||||
Amashanyarazi Yihuta Yigihe gito | 4.0 kV | ||||
Kubaga | 4.0kV(Uburyo rusange),2.0kV(Uburyo butandukanye) | ||||
Igihe cyo guhagarara nyuma yo gutangira | <3umunota | ||||
Ubushyuhe bwibidukikije | -25 ℃ - + 45 ℃ | ||||
Ubushuhe bugereranije | 40% -95% | ||||
Amashanyarazi akora | AC 55V - 600V | ||||
Gukoresha ingufu | <9W | ||||
Igipimo | 275mm * 196mm * 78mm | ||||
Ibiro | <1.5 kg |
1 - Kora kuri ecran
2 - Akabari
3 - Siba urufunguzo
4 - Sohora (inyuma) urufunguzo
5 - Mwandikisho
6 - Hindura urufunguzo
7 - Urufunguzo rw'umwanya
8 - Kugenda / Urufunguzo
9 - Amashanyarazi
10 - Guhindura amashanyarazi
11 - Ibyiciro bitatu bya voltage yinjiza sock (ABC)
12 - Icyambu cya USB
13 - Icyambu cyo gukosora RS232
14 - Kwishyuza sock
15 - Guhindura Bateri
17 - Ibyuma birindwi bifata ibyuma kuri A icyiciro cya clamp
18 - Ibikoresho birindwi bifata ibyuma bya B icyiciro cya clamp
19 - Ibikoresho birindwi bifata ibyuma bya C icyiciro cya clamp
20 - Imigaragarire yerekana ibimenyetso
21 - Icyiciro kigezweho cyinjiza & ibisohoka sock
22 - B icyiciro cyinjiza & ibisohoka sock
23 - C icyiciro cyinjiza & ibisohoka sock
24 - Kumanika indobo
Calibator yumurima kuri metero eshatu watt isaha | 1 set |
100A clamp Reba | Ibice 3 |
Impeta yo gukosora impeta | Igice 1 |
Umutwe w'amafoto df307 | Igice 1 |
USB disiki (software yo kuyobora) | Igice 1 |
Ibyiciro 3 byumurongo wumurongo mwinshi (harimo clamp na pin) | Igice 1 |
Icyiciro 3 cyubu (harimo clamp no gushiramo) | Igice 1 |
Icyiciro cya 3 cyumurongo 1A | Igice 1 |
Adapt | Igice 1 |
Umugozi w'amashanyarazi | Igice 1 |
Imishumi | Igice 1 |
Agasanduku k'ibikoresho | Igice 1 |
Igitabo cyibicuruzwa | Kopi 1 |
Icyemezo cyibicuruzwa | Kopi 1 |
Urutonde rwo gupakira uruganda | Kopi 1 |
Raporo y'ibizamini by'uruganda | Kopi 1 |