GDCL-V 20kV / 10kA Impulse Gukomatanya Umuhengeri Wibizamini bya Sisitemu Igisubizo
Ibikoresho bikurikije ibipimo byo guhuza ibizamini bisabwa kugirango sisitemu yo gukwirakwiza amashanyarazi make ya sisitemu ya SPD II, ifashe kubyara ingufu za impulse (1.2 / 50μs) hamwe na impulse ya (8 / 20μs), ikoreshwa mugupima icyiciro cya III no kugabanya ikizamini cya voltage ya SPD n'ibigize.Irakoreshwa kandi mugupima ibizamini bigezweho hamwe na voltage isigaye ya plaque ya MOV valve plaque, unit hamwe na surge ibikoresho birinda (SPD), cyangwa ikoreshwa mubindi bushakashatsi bwubushakashatsi.
Igikoresho gikoresha ecran ya tekinoroji na tekinoroji ya PLC, kandi imikorere nyamukuru yujuje ibikenerwa mu bicuruzwa byo mu gihugu ndetse n’amahanga SPD hamwe nibigize kugirango bigerageze guhuza ibizamini bitandukanye.
●GB 18802.1-2012 GB 18802.1-2012 Kubaga ibikoresho birinda bihujwe na sisitemu yo gukwirakwiza amashanyarazi make Igice cya 1: Ibisabwa nuburyo bwo gupima.
●GB / T 18802.12-2006 Kubaga ibikoresho birinda bihujwe na sisitemu yo gukwirakwiza amashanyarazi make Igice cya 12: Amabwiriza yo guhitamo no gukoresha.
●GB / T 16927.1-1997 Ubuhanga bwo gupima voltage Igice cya 1: Ibisabwa muri rusange.
●GB / T 16927.2-1997 Tekinike yo gupima voltage Igice cya 2: Sisitemu yo gupima.
●GB / T 17626.5-1999 Guhuza amashanyarazi ya elegitoroniki-Kugerageza no gupima tekinike-Kwipimisha ubudahangarwa bw'umubiri.
●IEC61000-4-5 Guhuza amashanyarazi (EMC) Igice cya 4-5: Uburyo bwo gupima no gupima - Kubaga ubudahangarwa bw'umubiri.
●IEC61643.11.
●IEC61643-21.
●Uburebure: 1000M.
●Ubushyuhe bwibidukikije: -5 ℃ ~ + 40 ℃, itandukaniro ntarengwa rya buri munsi: 25 ℃.
●Ubushuhe bugereranije: 85% (20 ℃) nta kondegene.
●Ibidukikije bya serivisi: Mu nzu.
●Nta mukungugu utwara, nta nkongi y'umuriro no guturika, nta byuma byangirika hamwe na gaze.
●Umuhengeri w'amashanyarazi atanga amashanyarazi ni sine, igipimo cyo kugoreka imiyoboro <5%, hamwe no guhindagurika kwa voltage <10%.
●Kurwanya impamvu: ≤0.5Ω.
Ibyingenzi byingenzi bya tekinike yikizamini ni:
●Imbaraga zinjiza: AC220V / 22kV / 3kVA, 60Hz
●Icyiza.DC yumuriro wa DC: DC12kV, max.kwishyuza 0.5A
●Igihe cyo kwishyuza: 22kV yarangiye muri 40 ~ 60s
●Gushiraho amashanyarazi ya voltage: Gutangiza ikizamini, kubanziriza gushiraho no guhindura voltage yumuriro
●Urwego rwa voltage isigaye: ≤2.5kV
●Guhindura polarite yikora: umubare wibyiza nibibi bya polarite impulse irashobora gushyirwaho ukwayo
●Inguni yicyiciro: 0-360 ° irashobora guhinduka
●Shira imbaraga zitanga uruziga
Imiravyo yumuriro 8 / 20μs, umurabyo ugereranije amplitude: 2-20kA
Ubushobozi: 16uF, voltage yumuriro: 12kV
Ubushobozi bwo gusohoka: 10% ~ 100%, anti-peak <20% (ntugahuze umutwaro wo gukuramo)
Imbogamizi zisohoka: 0.5Ω (10kV / 20kA)
●Gukomatanya umurongo utanga uruziga
Imiravyo yumuriro 8 / 20μs, umurabyo ugereranije amplitude: 1-10kA
Umuvuduko wumurabyo 1.2 / 50μs, umurabyo wumurabyo wapimwe amplitude: 2-20kV
Ubushobozi: 8uF, kwishyuza voltage 22kV
Ubushobozi bwo gusohoka: 10% ~ 100%, anti-peak <20% (ntugahuze umutwaro wo gukuramo)
Impedance yo gusohoka: 2Ω ± 10%
●Ibipimo by'inama y'abaminisitiri: 600 (L) * 800 (W) * 1800 (H) mm
●Ikizamini cyibikoresho: 400 (L) * 400 (W) * 300 (H) mm
●Uburemere: hafi 650 kg
●Inama y'Abaminisitiri ifite ibyuma byimukanwa hamwe n’inkunga ihamye, ishobora kwimurwa no guhagarara byoroshye
Igikoresho gifite mudasobwa ya Lenovo yose-imwe-imwe, ishyirwa iruhande rwigikoresho, kandi umurongo witumanaho uhujwe na mudasobwa kugirango isesengure rya flake
6.1
Impulse ubushobozi: 16uF * 2 / 12kV;
Ubushobozi bwa capacitor ikora kugeza kuri 22kV;
Umubare: 1;
6.2 Igice cyo Guhindura Umuhengeri
Ibice byo guhindura imirongo ya buri buryo bwakazi byahujwe ku kibaho gikingira ibintu byoroshye gusenya, gusimbuza no kubungabunga.
Uburyo butandukanye bwakazi burashobora guhindurwa byoroshye, kandi bikamenyekana na sensor ya fotoelektrike, ihita imenya imiterere yinjijwe, hanyuma ikayerekana kuri ecran ya ecran.
6.3
Koresha MDO3012 oscilloscope (cyangwa izindi moderi zitangwa numukiriya) zakozwe na Tektronix, icyitegererezo cya 100MHz, uburebure bwa 10k, hamwe na 8bit ihagaritse.Irashobora kuvugana na mudasobwa ikoresheje Ethernet.Amakuru arasesengurwa kandi atunganywa na PC, hanyuma agahita apimwa akanasesengurwa.Porogaramu irashobora kubika imiterere na data mugihe nyacyo.
6.4 Igice cyo kugenzura
Mugukoraho ecran na Mitsubishi PLC ikora sisitemu ya kabiri yo kugenzura ifite imiterere yoroshye, imikorere yoroshye no kuyitaho byoroshye.Igenzura ryimikorere ikoresha software yumwuga man-mashini, ikaba ari inshuti kandi yoroshye gukora.Ukoresheje ecran-10 ya ecran ya ecran, interineti yinshuti, byoroshye gukora no kwaguka.
6.5 Igikoresho cyo gupima
a) Igikoresho gipima voltage igikoresho gikoresha voltage ndende (RP1300H 300MHz 2.5kV) hamwe na voltage igabanya.Umuvuduko mwinshi wa voltage urashobora guhura na voltage isigaye ipima munsi ya 2.5kV.Ibyiza byujuje ibyifuzo byabakiriya kubisabwa kugirango voltage isigaye yibicuruzwa byerekana ibimenyetso.Igabanywa rya voltage rikoreshwa mubisabwa bya voltage bisigaye hejuru ya 2.5kV.
● Ibipimo byubushakashatsi:
Igihe cyo kuzamuka: 1.2nS
Kwinjiza inzitizi: 100MΩ
Umubare ntarengwa winjiza: 2.5kV (pulse)
Ukuri: 1%
●Ibice bitandukanya:
Umuvuduko ntarengwa winjiza: 2-20kV
Birakwiye kuri: 1.2 / 50
b) Igikoresho kigezweho cyo gupima gishobora gukoreshwa mugupima ubushobozi bwahagaritswe ukoresheje ingirakamaro-binyuze mubwoko bwa Rogowski.
● Ibipimo by'ibiceri:
Ikigereranyo cyo gupima ubu: 0.5-15kA
Ibisohoka ntarengwa ntarengwa: 300V
Ikigereranyo cyo gutandukana: 0.02V / A.
●Igice kinini cya voltage yibikoresho cyitaruye kandi gihagaze.
●Igenzura ryinama ryashizweho, kandi impera yumuzingi wikizamini ihora ihagaze.
●Itara ryerekana umutuku mwinshi cyane ryerekana ko umuyaga mwinshi ukora, naho urumuri rwicyatsi kibisi rwinshi rwerekana ko nta kaga gakomeye gafite.
●Iyo hafunguwe urugi rwo gukingira urugi cyangwa ibintu byihutirwa byafunguwe, umuyaga mwinshi urekurwa hasi, ibikoresho ntibikora, kandi itara ryerekana interineti ryaka.
●Inzira nyamukuru yibikoresho ifite insinga ya fuse yo kurinda birenze urugero kugirango umutekano ubeho.
●Ibikoresho byose byanditseho ikimenyetso cyumutekano.
●Ikizamini kirashobora guhagarikwa / gukanda na bouton yihutirwa yo guhagarara, hamwe nibice byimbere byumuvuduko mwinshi hamwe nibisohoka bigufi-bizunguruka bishobora kugaragara.
●Bifite ibikoresho byintoki byo gufata neza abaminisitiri.
●Ibikoresho bifite ibikoresho bya Omron electromagnetic bifunga kumuryango wibicuruzwa, kandi ntibyemewe gukingura urugi rwibicuruzwa mugihe cyibizamini kugirango umutekano urusheho kuba mwiza.
Izina | Icyitegererezo | Qty. |
CD ya software |
| 1 |
Fuse | 3A 10 * 38 | 1 |
Amashanyarazi | 35W-24V | 1 |
Umuyoboro mwinshi wa diode | 50kV 0.5A | 2 |
Inductance nkeya irwanyaor | 1 Ω / 100W | 4 |
BNC Umuyoboro wa kabili | 75-5 f9 | 1 |
Oya. | Izina | Icyitegererezo / Ibisobanuro | Qty. | Ongera wibuke |
1 | Imashini itanga amashanyarazi | GDCL-V 20kV / 10kA | 1 | 20kV / 10kA |
2 | Sisitemu yo gupima | GDCL-V 20kV / 10kA | 1 | Hatabayeho kugenzura abaminisitiri |
3 | Umuhengeri1 | 8 / 20μs | 1 | |
4 | Umuhengeri2 | 1.2 / 50 8/20 | 1 | |
5 | Inkoni y'ubutaka | JDB5 | 1 | |
7 | Inyandiko | Igitabo, raporo yikizamini no gushushanya amashanyarazis | 1 | |
8 | Igikorwaumutwaro w'amashanyarazi | 0-500VAC | 1 | |
9 | Uburyo bwo guhuza | 9Uf * 2 | 2 |